What are some common malfunctions of automatic dripping machines?

Jan 07, 2026 Leave a message

I. Power Supply and Circuit Malfunctions

Symptom: Indicator light does not respond after startup.

Cause: Damaged wiring or circuit board; interrupted connection between gear shaft and piston push shaft.

Solution: Check circuit board; reinstall or replace piston push shaft.

II. Electrode and Measurement Malfunctions

Symptom: Millivoltmeter pointer swings erratically.

Cause: Loose electrode wiring; contaminated reference electrode solution; or circuit interruption.

Solution: Check wiring; replace electrode or clean reference electrode.

III. Titrate and Piping Malfunctions

Symptom: Titrate cannot be injected or flow is blocked.

Cause: ... Causes: Burette leakage, loose connection of suction tubing or infusion tubing

Solution: Check burette sealing ring, tighten connection nut

IV. Mechanical Component Failure

Symptom: Interruption in connection between titrator gear shaft and piston push shaft

Cause: Loose threaded interface or shaft connection failure

Solution: Reinstall connection or replace piston push shaft

V. Other Common Problems

Air Bubble Alarm: Air bubbles in tubing or empty solution bottle

Low Battery Voltage Alarm: Insufficient battery power

Drop Count Alarm: Empty infusion bottle, flow rate regulator not turned on, or sensor damaged