I. Potential/Measurement Issues: Unstable and fluctuating potential readings are often caused by electrode aging and contamination, external electromagnetic interference, or poor grounding. Large deviations between measured and theoretical pH values are frequently due to expired or contaminated buffer solutions, improper electrode calibration, or electrode aging.
II. Titration System Mechanical Issues: Leaks in the burette or uneven dispensing are often caused by loose tubing connections, aging seals, or blockages from deposits in the tubing. Air bubbles in the titration circuit directly affect dispensing accuracy, often caused by loose tubing connection nuts.
III. Post-Calibration Accuracy Issues: Inaccurate burette coefficient calibration directly leads to deviations in dispensing volume. Inadequate tubing cleaning (failure to clean with standard solution at least 6 times) and improper jump point settings will result in inaccurate measurements even after calibration.
IV. Solenoid Valve Control Issues: A flashing titration lamp without dispensing is often due to improper adjustment of the solenoid valve support screw. Leaks even after the solenoid valve is closed are often caused by aging, deformation, or decreased elasticity of the rubber tubing.
V. Counter-related issues: If the counter cannot return to zero after calibration and replenishment, it is often due to loose screws on the counter shaft or misaligned gears. Manual zeroing and re-fixing of the gears are required to reset the counter.